HUMY 301 / DYNAmoist

Inline moisture measurement offered in the MÜTEC and DYNA portfolios. Monitor moisture in bulk solids during production; assess installation, product contact and material-specific calibration together.

Find the right moisture measurement for your application

Share your material and process details. We can complete any unknown technical details together.

Moisture measurement application form →

Inline moisture measurement tracks changes in the moisture of flowing bulk solids during production. Reliable results require continuous and sufficient material contact with the sensor, stable flow conditions, and product-specific calibration against laboratory samples taken from the same point.

Quick response

  • The sensor face must be continuously covered by enough material.
  • Factory settings alone are insufficient; calibrate for the specific product and process.
  • Take the laboratory sample at the same time and point as the material seen by the sensor.
  • Different products and recipes may need separate calibration curves.
  • The choice between HUMY 301 and HUMY 3000 depends on sensor, logging and control requirements as well as the need for a display.
The sensor signal is converted to a moisture value using laboratory samples taken at the same time from the same product.

How does inline moisture measurement work?

The HUMY 301 sensor detects moisture-dependent changes in the electrical properties of the material in front of it. Converting the raw signal to a moisture percentage requires a calibration curve for that product. There is no universal calibration for different materials.

Material height, bulk density, air gaps and coverage of the sensor face can affect the measurement. Establishing stable process conditions comes before adjusting filters.

What is the difference between HUMY 301 and HUMY 3000?

MÜTEC HUMY 301 inline moisture system
PLC / software integration

HUMY 301 / DYNAmoist

Suitable when the moisture value can be displayed through the PLC, control system or an appropriate plant display.

You are currently viewing this product
MÜTEC HUMY 3000 inline moisture system with display
Option with a local display

HUMY 3000

Suitable when the plant has no PLC or display for moisture values and a separate display unit is needed beside the sensor.

HUMY 3000 product details
Selection criterionHUMY 301 / DYNAmoistHUMY 3000
Moisture displayPLC or existing operator displaySeparate display unit beside the sensor
Reason for selectionExisting plant display infrastructureLocal display required
Measurement taskInline moisture with product-specific calibrationInline moisture with product-specific calibration
Automation connectionAnalog output or Modbus RTU; matched to device/software versionAnalog output, alarms and RS485/Modbus; matched to system configuration
Product changeProduct-specific curve and correct selection requiredManufacturer specifies 24 product/product-group parameter sets
Logging requirementPlanned with PC software and connectionsAssess integrated logging and local monitoring needs
Final selectionSensor, process connection, signals and software versionSensor, display, mounting and system scope

Recipe count, sensor count and selection input differ

The number of stored calibrations is not the number of sensors that can be connected simultaneously. The HUMY 301 brochure describes 24 product calibrations and selection of 16 products via digital inputs, while the current manufacturer webpage specifies selection of 8 data sets. Use the hardware and software revision of the installed or quoted device. The 24 product groups specified for HUMY 3000 are not a sensor count either.

In the quotation, state the number of measuring points and recipes, whether the recipe is selected by the PLC or operator, and where historical moisture data will be stored.

Let us review your moisture measuring point together

After comparing HUMY 301 and HUMY 3000, share material and installation details in the short form.

Where should an inline moisture sensor be installed?

A good mounting point maintains a stable and sufficient material depth over the sensor face. Material movement should be as consistent as possible; limit air gaps, irregular filling and stationary material buildup in front of the sensor.

Good measuring conditions

  • Continuous material coverage
  • Stable material movement
  • Sufficient material depth
  • Sampling point near the sensor
  • Access for maintenance and cleaning

Conditions that complicate measurement

  • Partly filled channel
  • Variable material depth
  • Air gaps
  • Dead zone in front of the sensor
  • Sampling at a different point
HUMY 3000 brochure installation examples for hoppers, screws, free fall and belts
HUMY 3000 manufacturer brochure, p. 2. These are example layouts and must not be used as dimensional drawings for HUMY 301. Open brochure (PDF) →
Measuring pointRequired conditionSite check
Belt / chuteStable product layer on the sensorLayer depth, product separation from sensor and sampling access
Screw conveyorRepresentative product contact each revolutionFill, blade passage and filter; verify layout by application
Vessel / hopper outletRenewed product in front of the sensorDead zones, bridging and surface buildup
Intermittent processDefined product contact time for measurementExclude empty periods and coordinate sampling

Installation on a screw conveyor

Illustrative inline moisture sensor point in a screw conveyor

The manufacturer's fish-feed application report recommends offsetting the sensor by approximately 20° in the screw's direction of rotation. Avoid mounting it directly underneath: the screw flight leaves a gap at the bottom, where material movement may be unstable.

The same report specifies sufficient screw fill, a calibration period of approximately 30–60 seconds and a filter period of at least 30 seconds on the measurement output. These values relate to that application; determine settings for another process by observing its material movement.

How is product-specific calibration performed?

01

Define the reference method

Set the oven, halogen moisture analyser or other method and its operating procedure.

02

Take simultaneous samples

Record the sensor's raw value and take a sample close to the sensor; transport it in a sealed container.

03

Cover the moisture range

Collect data at different moisture points representative of the process range, not just around the normal value.

04

Create the curve

Determine the slope and offset relating sensor signals to laboratory results.

05

Perform an independent check

Check the result under different operating conditions using new paired samples that were not used for calibration.

The manufacturer's application report recommends calibrating during operation, evaluating at least two moisture values and analysing at least three laboratory samples at each point. More paired samples across a wider moisture range help the regression represent the process range.

Why does the sensor disagree with the laboratory result?

Source of deviationEffectCheck
Sample taken at a different timeSensor and lab compare different productSimultaneous sample
Sample waits in an open containerHot product especially may lose moistureSealed sample container
Laboratory method changesReference result may changeOne method and fixed procedure
Product depth changesMaterial volume seen by the sensor changesStable, sufficient coverage
Flow conditions changeContact and measurement behavior may changeStable process conditions
Air gap formsMeasurement is not sufficiently representativeCorrect mounting location
Recipe changesCalibration relationship may changeProduct-specific curve

Review trends and sampling procedures before individual readings. Do not expect closer agreement from the sensor without first assessing the repeatability of the laboratory reference method.

Can one sensor measure several products?

One sensor can be used for different products, but each product or substantially different recipe group may require its own calibration curve. When the product changes, the correct curve must be selected reliably by an operator, digital input or plant control system.

Define the product list and how changes are communicated during quotation. This prevents many later complaints that the sensor is reading incorrectly.

When is recalibration necessary?

Raw material or supplier changeRecipe or grain-structure changeChange of mounting pointChange in product depth or flowChange of laboratory methodSensor-surface buildup or wearPersistent difference between measurement and process result

Rather than scheduling calibration solely by date based on an assumption of electronic drift, monitor periodic check samples and process changes.

What do the PT1, Median and Screw filters do?

PT1

Smooths short-term fluctuations. A longer filter time can produce a steadier reading but slow the response to process changes.

Median

Helps prevent sudden outlier readings from unduly affecting the displayed value.

Screw

Helps handle periodic material contact and signal variations caused by screw rotation.

A filter cannot correct insufficient material coverage or a poor mounting point. Set the final filter and parameters using process data.

For which materials can inline moisture measurement be considered?

Feed and pelletsGrain and flourStarchNuts and granular foodsSand and mineralsChemical powders and granulesPlastic granulesSawdust and chips

This list does not mean every material can be measured with the same configuration. Assess product samples, target moisture range, mounting conditions and laboratory method separately for each application.

Common mistakes

Insufficient product depth

Measurement conditions change when coverage is not continuous and uniform.

Narrow-range calibration

One or closely spaced moisture values do not represent the operating range.

Wrong sampling time

Earlier or later product is compared with the laboratory instead of the measured product.

Open-container sample transport

Hot samples may lose moisture before reaching the lab.

One curve for all products

Recipe/material changes may require separate calibration.

Trying to fix mounting with filtering

Smoothing the signal does not remove air gaps or irregular contact.

Promising accuracy with one number

Assess results with calibration, reference method and mounting quality.

No defined acceptance procedure

Define sampling, reference laboratory and acceptance criteria before commissioning.

Applications reviewed by Infomak

Illustrative inline moisture installation on a chain conveyor in feed processing

HUMY 301 demonstration in a feed plant

On a chain conveyor, material coverage, measurement stability and process response were reviewed. PT1, Median and Screw filters were considered against the product flow. The company name and unverified performance figures are withheld.

Belt conveyors carrying sand and gravel at a hopper outletSource: MÜTEC sand moisture measurement application report, 05/04/2024.

Inline moisture measurement in sand feed

The photograph shows belt conveyors carrying sand and gravel from the hoppers. The report examines monitoring inlet moisture under changing weather conditions and keeping material density and contact stable at the measuring point.

Frequently asked questions

Does an inline moisture sensor replace the laboratory?

It provides continuous readings and trends during production. A defined laboratory reference remains necessary for initial product calibration and later checks.

Why does the sensor disagree with the laboratory result?

Differences often arise because sensor and laboratory compare different material. Sample the sensor's material simultaneously, transport it sealed and use the calibration laboratory method.

Does each product need separate calibration?

Recipe or physical changes can alter the moisture-to-signal relationship. Consider separate curves for different products or substantial recipe changes.

How often should the sensor be recalibrated?

Track process changes rather than a fixed calendar. Check calibration after changes to raw material, recipe, supplier, location, product depth or laboratory method.

Should I select HUMY 301 or HUMY 3000?

Assess HUMY 301 with existing PLC/display/software and HUMY 3000 for a local display unit. Confirm sensor, mounting, recipes, logging and communications for final selection.

Where is the best inline moisture mounting point?

Prefer stable, sufficient product coverage, steady movement and nearby sampling. Screws, hoppers, chutes or belts may suit; determine exact placement by observing flow.

Related products and technical information

Manufacturer documents

HUMY 301 / DYNAmoist

HUMY 301 comprises an FMS 410 moisture sensor and a DIN-rail evaluation unit. The sensor measures product in the process; the unit makes moisture data available for display, alarms and automation.

Tracks moisture changes on the production line while laboratory results are pending. It can provide feedback for drying or humidification control; the control strategy is designed with the plant automation.

How does capacitive moisture measurement work?

Material in front of the sensor forms the dielectric medium of an open capacitor. Its electrical behavior changes with water content. A product-specific calibration curve converts this change into moisture.

A single universal calibration therefore cannot cover different raw materials. Changes in product recipe and bulk density are included in measurement assessment.

For which processes is it considered?

Measuring pointPurpose
Dryer outletMonitor outlet moisture and provide feedback for dryer settings
Humidification / mixing lineMonitor the effect of added water and moisture changes
Screw conveyor outlet or product chuteMeasure flowing product with representative contact
Conveyor belt or vesselMonitor moisture with a consistent product layer in front of the sensor

Application conditions are assessed for feed, grain, sand, fertilizer, plastic granules, pellets and forest products. Measurement point and calibration are determined separately for each material.

Commissioning and calibration

First select a point where the sensor continuously contacts product. Keep velocity, layer depth and bulk density as stable as possible; eliminate air gaps and deposits in front of the sensor.

The manual recommends calibration at at least two moisture levels in the running process, preferably using up to five reference points. Take a sample near the sensor while recording its raw value, then measure it by a reference method.

After calibration, compare results under different operating conditions. Laboratory references remain important for calibration and periodic verification.

Product recipes, digital selection and software

The downloadable HUMY 301 brochure specifies storage of 24 calibrations, with 16 selectable via digital inputs. The current website lists 8 datasets for digital selection. Therefore, match the installed hardware and software version before quotation.

Displaying multiple measuring points with HUMIConfig is different from the number of recipes in one device. For existing devices, also check HUconfig version and connection compatibility.

Technical documents

HUMY 301 · User manual (English)Download document ↓HUMY 301 brochure linked by the manufacturer for DYNAmoist · ENDownload document ↓HUMY 301 · Sensor operating manual (EN · 16 pages)Download document ↓

When is it suitable?

For continuous product moisture information in drying, humidification or quality monitoring, provided suitable product contact and material-specific calibration are available.

Technical selection and configuration

Measuring principleCapacitive / product-specific calibration
SystemFMS 410 sensor + HUMY 301 evaluation unit
Transfer to automationFrom the evaluation unit: 0/4–20 mA or 0–10 V; RS485 Modbus
Product memoryManufacturer brochure: 24 calibrations; 16 products selectable via digital inputs
ParameterisationHUMIConfig or HUconfig appropriate to the device/software version
Unit power supply24 VDC (18–30 V); according to the operating manual
Sensor surfacePOM, ceramic or Teflon options according to material and wear conditions
Temperature and pressureConfirm according to sensor version, process connection and Ex selection

Confirm numeric limits and connection options against the current manufacturer datasheet and exact order code.

Installation notes

Position the sensor on a belt, screw outlet, chute or vessel to maintain regular product contact. Do not place it beside free-falling product like a non-contact probe; direct product over the sensor if necessary. Exact placement depends on material flow and manufacturer installation instructions.

ATEX and certificates

For certified versions, check the full device type code, marking and manufacturer documentation together.

MÜTEC official product information ↗

Does HUMY 301 directly provide moisture percentage?

After product-specific calibration, the measurement can be used as a moisture value. The raw sensor signal must not be interpreted as a universal moisture percentage without a material reference.

Does 0.1% accuracy apply to every product?

The brochure's accuracy of up to 0.1% depends on the product and application conditions. Expected performance is assessed using actual material and reference measurements.

Information for a quotation

Provide material name, target moisture range, temperature, capacity, a measurement-point photo or drawing, required output signal and any Ex zone.

Related solution area

This shared page provides the measuring principle, installation and calibration information for HUMY 301 / DYNAmoist. Quotations are based on the device nameplate, order code and hardware/software version.